A near-quantum-limited diamond maser amplifier operating at millikelvin temperatures
Abstract
Microwave quantum technologies require amplification of weak signals with minimal added noise at millikelvin temperatures. This stringent demand has been met with superconducting parametric amplifiers. While masers offer another fundamental approach, their dependence on cryogenic operation has historically posed challenges for classical communication technologies — a barrier that does not apply to microwave quantum technologies. In this work, we demonstrate an ultra-low-noise maser amplifier utilizing impurity spins in diamond at millikelvin temperatures. We achieve power gains exceeding , a minimum noise temperature of (corresponding to noise photons), and a maximum output compression point of at . Our results establish masers as viable components of microwave quantum technologies.
The tremendous developments in quantum information technologies that operate at millikelvin temperatures in the microwave frequency band over the past two decades have relied on ultra-low-noise amplification with near-quantum-limited noise [1, 2]. This demand has been met primarily by superconducting circuit technologies [3, 4, 5, 6]. Microwave amplification via masers (microwave amplification by stimulated emission of radiation) — the microwave-frequency counterpart of the laser — has been predicted to possess quantum-limited noise performance, particularly when thermal noise is absent[7] and sufficient population inversion is achieved [8, 9, 10].
Diamond masers, particularly those based on negatively charged nitrogen-vacancy () centers, have emerged as both oscillators [9, 11] and amplifiers at room temperature [10] and cryogenic temperatures [12]. In addition, cavity cooling [13, 14, 10] and enhanced magnetometry [15] at room temperature have been demonstrated, suggesting the potential of diamond-based masers as resources for microwave quantum technology applications. However, the requirement for optical pumping to establish population inversion in centers presents a significant challenge for applications at millikelvin temperatures due to the high laser powers required and the limited cooling power of dilution refrigerators. In this work, we achieve population inversion in substitutional nitrogen defects (P1 centers) in diamond through microwave pumping at power levels of nanowatts or less. We demonstrate amplification with near-quantum-limited noise at millikelvin temperatures.
Maser amplifier device and setup

Fig. 1A illustrates the maser amplifier device. It consists of a loop-gap microwave resonator [16] that houses a diamond crystal containing approximately of P1 centers, together with about of negatively-charged nitrogen-vacancy (NV-) centers [17, 16], with dimensions of . The resonator’s frequency is and its internal quality factor is . In this work, we operate the resonator with two different external quality factors, and , by adjusting the number of turns of the coupling loop antenna.
Fig. 1B illustrates a simplified measurement setup inside the refrigerator. The maser amplifier device is thermalized to the mixing chamber plate at in a dilution refrigerator. Two microwave input lines, one heavily attenuated for the probe and another moderately attenuated for the pump, are combined through a directional coupler. After being combined, the pump and probe microwave tones go to the maser amplifier device through a circulator, which in turn directs the reflected signal to the measurement line and decouples the thermal noise from the following HEMT (high electron mobility transistor) amplifier placed at the stage. An impedance-matched temperature-variable noise source [18] is integrated along the probe line, which is weakly thermalized to the cold plate at approximately . We use this noise source to characterize the noise temperature of our diamond maser amplifier.
Operating principle

The electron spins of the P1 centers are the active medium in our maser amplifier. These centers possess an electron spin and a nuclear spin , with a hyperfine interaction tensor characterized by and . Under a static magnetic field aligned parallel to the cubic crystallographic axis of the diamond, the electron spin states () split into three levels depending on the nuclear spin states, P1+, P10, and P1-, as shown in Fig. 2A. The corresponding transition frequencies, denoted as , , and , satisfy an energy conserving relationship , which facilitates a four-spin cross-relaxation process between the central transition () and the two satellite transitions () [19, 20, 21].
The operating principle of our maser amplifier is described in Figs. 2B. First, we apply a static magnetic field to align the frequency with the resonator frequency , and then pump the P10 transition at with a rate , as illustrated in Fig. 2B (1). This pumping triggers the excitation of one P1+ and one P1- spin by exchanging the energy with two P10 spins through the four-spin cross-relaxation process [19, 20] at a rate of , as depicted in Fig. 2B (2). Ordinarily, such pumping would lead to saturation across all transitions. However, the dynamics are modified by the difference between the relaxation times of the spins in P1- and P1+, as demonstrated in Fig. 2C, with respective values of and . The faster relaxation of P1- is induced by another “fast-relaxing” spin system, which coexists within the diamond crystal and exhibits relaxation times on the order of ten milliseconds (see inset of Fig. 2C and [22]). The transition frequencies of these “fast-relaxing” spins largely overlap with the P1- transition around , thereby accelerating the relaxation of the P1- spins through magnetic dipole-dipole interactions, as indicated by in Fig. 2B (3). This imbalance in relaxation times prevents saturation and maintains continuous excitation of the P1+ state, resulting in population inversion. Although the exact nature of these fast-relaxing spins remains unidentified, they might be either neutrally-charged nitrogen-vacancy (NV0) centers [23, 24, 22] or nickel centers [25]. We refer to them as “fast-relaxing spins” throughout this paper.
Gain and bandwidth

Fig. 3A shows the time evolution of the inverted spin population between the upper () and lower () levels of P1+ transition, extracted from the measured reflection coefficients [22]. Above a pump power of at the device input, becomes positive, indicating population inversion. We also simulate the dynamics using rate equations that include all P1 states and their interactions [22], which are shown by the solid lines in Fig. 3A.
Fig. 3B displays the maser’s gain spectra under several different pump powers measured by a vector network analyzer (VNA) under a static magnetic field of . Although the spectra show negative gains at weaker pump powers until , the gain becomes positive and increases above . With a pump of , the gain exceeds with a bandwidth of approximately , which results in the gain-bandwidth product of . Fig. 3C plots the gain-bandwidth products under various pump powers for the two different external quality factors, and , of the microwave resonator, as a function of the inverted spin population . The inset presents the gain. For the maser amplifier device with , the system enters the self-oscillation regime at , where coherent microwave radiation is spontaneously emitted. This is visible as the saturation of above the gain-bandwidth product of approximately , as also indicated by the red dot-dashed lines in the insets of Fig. 3C and A.
Next, we demonstrate the frequency tunability of the diamond maser’s gain profile by varying the external static magnetic field . As the transition frequencies of the P1 centers shift with , the maser’s gain profile can also be tuned accordingly. Fig. 3D shows the measured maser gain versus , demonstrating tunability over in the central gain and about in the full-width at half-maximum.
Noise temperature

To assess the noise performance of our maser amplifier, we measured the noise power spectrum in the presence of a weak microwave probe tone as a reference, with the maser turned on and off. The results displayed in Fig. 4A show a signal-to-noise ratio improvement of approximately . This indicates that our maser amplifier achieves a noise level approximately four times lower than that of the following measurement line, which is dominated by the noise level of the HEMT amplifier.
To quantitatively evaluate our maser amplifier’s noise temperature, we use the impedance-matched temperature-variable noise source [18]. This device enables independent control of the input noise temperature to the maser amplifier, resulting in accurate determination of noise temperature. It is important to note that we also account for the attenuation of the lines between the noise source and the maser amplifier [22], as well as the insertion losses at the directional coupler and the circulator. We have independently measured these attenuations to obtain the accurate value of (Supplementary Text).
First, we characterized the base noise temperature of our measurement line with the maser amplifier turned off. In this configuration, the noise power after the HEMT amplifier as a function of , detected within a bandwidth of , is given by [18]
(1) |
where and are the gain and noise temperature of the HEMT amplifier, respectively, and is the Boltzmann constant. The results, shown in the inset of Fig. 4B, indicate that .
Subsequently, we turned on the diamond maser amplifier and repeated the noise temperature measurements at various maser amplifier gains. In this configuration, the output noise power after the HEMT amplifier is given by
(2) |
Here, and represent the gain and noise temperature of the maser amplifier, respectively. As presented in Fig. 4B, the minimum noise temperature of the entire system is approximately . After subtracting the contribution from , the minimum noise temperature of the diamond maser amplifier is obtained to be , corresponding to approximate noise photons of .
The measured noise temperature , which is nearly three times higher than the quantum limit of , corresponding to approximately in our setup, primarily results from imperfect population inversion[8, 10]. While the spins in the higher energy level amplify the incoming signal through stimulated emission, those remaining in the lower energy level absorb photons. This counteraction effectively introduces additional noise photons, quantified by , where negative spin temperature and the inversion ratio are linked through
(3) |
with denoting the reduced Planck constant. As shown in Fig. 3A, the measured at gain corresponds to a noise temperature of [10], in good agreement with the measured . It is important to note that the measured values are by no means the fundamental limits in our system; they are constrained by the slow relaxation rate of the spins, , which in turn limits the effective pump rate on the P1+ spins. With a slightly faster of , which could be realized by engineering a higher density of the fast-relaxing spins, the noise temperature could be reduced to . Additionally, improving the internal quality factor would further lower the maser amplifier’s noise temperature. For instance, an order-of-magnitude improvement in to , achievable without reducing the coupling constant by employing a dielectric resonator [11, 26, 27], could lower the noise temperature by approximately .
Compression point
To assess the dynamic range of our maser amplifier, we measured the compression points at various maser amplifier gains. Fig. 4C shows the amplifier gains as a function of input probe power. The output (input) compression points () are measured to be approximately () at a gain of , and approximately () at a gain of . These measured compression points are to higher than those typically reported in Josephson traveling-wave parametric amplifiers [4, 2], but are two or three orders of magnitude lower than those achieved by state-of-the-art kinetic inductance traveling-wave parametric amplifiers [28, 29].
Conclusion
We demonstrated a diamond maser amplifier operating at millikelvin temperatures and achieved low noise photon levels of . The noise level could be further reduced by increasing the effective pump rate through a higher density of fast-relaxing spins and by employing a resonator with a higher internal quality factor. By revisiting masers at millikelvin temperatures, our work reestablishes maser technology as a robust and low-noise alternative for microwave amplification in quantum technologies, such as quantum computing, magnetic resonance spectroscopy, and particle physics.
Acknowledgements.
We thank J. Pla, Ç.O. Girit, P. Bertet, A. Bienfait, D. Xiao, R.-B. Liu, J.J.L. Morton, K. Mølmer, E. Flurin, M. Hatifi, F. Quijandría, J. Teufel, Y. Ota, S. Iwamoto, and S. Kono for their discussions, and M. Ishida for assistance with device illustration. We are also grateful to the members of the Hybrid Quantum Device Team within the Science and Technology Group and the Experimental Quantum Information Physics Unit at Okinawa Institute of Science and Technology (OIST) Graduate University for their insights. Additionally, we acknowledge the support from S. Ikemiyagi, M. Kuroda, and P. Kennedy at the Engineering Section of OIST.Funding:
This work was supported by the Proof-of-Concept (PoC) program by OIST Innovation, the JST Moonshot R&D Program (Grant No. JPMJMS2066), JST-PRESTO (Grant No. JPMJPR15P7), JSPS KAKENHI Grant-in-Aid for Scientific Research (B) (Grant No. 18H01817), Grant-in-Aid for Scientific Research on Innovative Areas (Grant No. 18H04295), the Science Research Promotion Fund from the Promotion and Mutual Aid Corporation for Private Schools of Japan (PMAC), and the Research Encouragement Grant from the AGC Inc. Research Foundation. M.O. and T.H. acknowledge financial support from the JSPS Grant-in-Aid for Fellows (Grant Nos. 23KJ2135 and 24KJ2178), and J.R.B. acknowledges support from OIST Graduate University.
Author contributions:
M.O. and C.-P.L. contributed equally to this work and were primarily responsible for data collection with partial assistance from V.P.M.S., T.H., I.K., and Y.Ku. M.O. and C.-P.L. analyzed the data with assistance from T.H., I.K., V.P.M.S., and Y.Ku. Y.Ku. conceived the project and initiated the preliminary experiments with J.R.B. and P.M. S.O., T.O., Y.Ko., and J.I. were responsible for the preparation of the diamond sample. Y.Ku. and H.T. jointly supervised the project. All authors contributed to the manuscript.
Competing interests:
There are no competing interests to declare.
Data and materials availability:
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- Clerk et al. [2010] A. A. Clerk, M. H. Devoret, S. M. Girvin, F. Marquardt, and R. J. Schoelkopf, Introduction to quantum noise, measurement, and amplification, Reviews of Modern Physics 82, 1155 (2010).
- Esposito et al. [2021] M. Esposito, A. Ranadive, L. Planat, and N. Roch, Perspective on traveling wave microwave parametric amplifiers, Applied Physics Letters 119, 120501 (2021).
- Yamamoto et al. [2008] T. Yamamoto, K. Inomata, M. Watanabe, K. Matsuba, T. Miyazaki, W. D. Oliver, Y. Nakamura, and J. S. Tsai, Flux-driven Josephson parametric amplifier, Applied Physics Letters 93, 042510 (2008).
- Macklin et al. [2015] C. Macklin, K. O’Brien, D. Hover, M. E. Schwartz, V. Bolkhovsky, X. Zhang, W. D. Oliver, and I. Siddiqi, A near–quantum-limited Josephson traveling-wave parametric amplifier, Science 350, 307 (2015).
- Eom et al. [2012] B. H. Eom, P. K. Day, H. G. LeDuc, and J. Zmuidzinas, A wideband, low-noise superconducting amplifier with high dynamic range, Nature Physics 8, 623 (2012).
- Malnou et al. [2021] M. Malnou, M. Vissers, J. Wheeler, J. Aumentado, J. Hubmayr, J. Ullom, and J. Gao, Three-Wave Mixing Kinetic Inductance Traveling-Wave Amplifier with Near-Quantum-Limited Noise Performance, PRX Quantum 2, 010302 (2021).
- Weber [1957] J. Weber, Maser Noise Considerations, Physical Review 108, 537 (1957).
- Siegman [1964] A. E. Siegman, Microwave solid-state masers (McGraw-Hill, 1964).
- Jin et al. [2015] L. Jin, M. Pfender, N. Aslam, P. Neumann, S. Yang, J. Wrachtrup, and R.-B. Liu, Proposal for a room-temperature diamond maser, Nature Communications 6, 1 (2015).
- Day et al. [2024] T. Day, M. Isarov, W. J. Pappas, B. C. Johnson, H. Abe, T. Ohshima, D. R. McCamey, A. Laucht, and J. J. Pla, Room-Temperature Solid-State Maser Amplifier, Physical Review X 14, 041066 (2024), publisher: American Physical Society.
- Breeze et al. [2018] J. D. Breeze, E. Salvadori, J. Sathian, N. M. Alford, and C. W. M. Kay, Continuous-wave room-temperature diamond maser, Nature 555, 493 (2018).
- Sherman et al. [2022] A. Sherman, O. Zgadzai, B. Koren, I. Peretz, E. Laster, and A. Blank, Diamond-based microwave quantum amplifier, Science Advances 8, eade6527 (2022), publisher: American Association for the Advancement of Science.
- Ng et al. [2021] W. Ng, H. Wu, and M. Oxborrow, Quasi-continuous cooling of a microwave mode on a benchtop using hyperpolarized NV- diamond, Applied Physics Letters 119, 234001 (2021), publisher: American Institute of Physics.
- Fahey et al. [2023] D. P. Fahey, K. Jacobs, M. J. Turner, H. Choi, J. E. Hoffman, D. Englund, and M. E. Trusheim, Steady-State Microwave Mode Cooling with a Diamond N-$V$ Ensemble, Physical Review Applied 20, 014033 (2023), publisher: American Physical Society.
- Wu et al. [2022] H. Wu, S. Yang, M. Oxborrow, M. Jiang, Q. Zhao, D. Budker, B. Zhang, and J. Du, Enhanced quantum sensing with room-temperature solid-state masers, Science Advances 8, eade1613 (2022), publisher: American Association for the Advancement of Science.
- Ball et al. [2018] J. R. Ball, Y. Yamashiro, H. Sumiya, S. Onoda, T. Ohshima, J. Isoya, D. Konstantinov, and Y. Kubo, Loop-gap microwave resonator for hybrid quantum systems, Applied Physics Letters 112, 204102 (2018).
- Grezes et al. [2014] C. Grezes, B. Julsgaard, Y. Kubo, M. Stern, T. Umeda, J. Isoya, H. Sumiya, H. Abe, S. Onoda, T. Ohshima, V. Jacques, J. Esteve, D. Vion, D. Esteve, K. Mølmer, and P. Bertet, Multimode Storage and Retrieval of Microwave Fields in a Spin Ensemble, Physical Review X 4, 021049 (2014).
- Simbierowicz et al. [2021] S. Simbierowicz, V. Vesterinen, J. Milem, A. Lintunen, M. Oksanen, L. Roschier, L. Grönberg, J. Hassel, D. Gunnarsson, and R. E. Lake, Characterizing cryogenic amplifiers with a matched temperature-variable noise source, Review of Scientific Instruments 92, 034708 (2021), tex.ids= simbierowiczCharacterizingCryogenicAmplifiers2021a publisher: American Institute of Physics.
- Sorokin et al. [1960] P. P. Sorokin, G. J. Lasher, and I. L. Gelles, Cross Relaxation Studies in Diamond, Physical Review 118, 939 (1960).
- Ma et al. [2019] C. Ma, Z. Wang, Y. Wu, Z. Bao, Y. Song, H. Zhang, and L. Duan, Four-spin cross relaxation in a hybrid quantum device, Physical Review A 100, 012322 (2019).
- Zhang et al. [2021] G.-Q. Zhang, Z. Chen, D. Xu, N. Shammah, M. Liao, T.-F. Li, L. Tong, S.-Y. Zhu, F. Nori, and J. Q. You, Exceptional Point and Cross-Relaxation Effect in a Hybrid Quantum System, PRX Quantum 2, 020307 (2021), publisher: American Physical Society.
- [22] Materials and methods are available as supplementary materials.
- Baier et al. [2020] S. Baier, C. Bradley, T. Middelburg, V. Dobrovitski, T. Taminiau, and R. Hanson, Orbital and Spin Dynamics of Single Neutrally-Charged Nitrogen-Vacancy Centers in Diamond, Physical Review Letters 125, 193601 (2020), publisher: American Physical Society.
- Barson et al. [2019] M. S. J. Barson, E. Krausz, N. B. Manson, and M. W. Doherty, The fine structure of the neutral nitrogen-vacancy center in diamond, Nanophotonics 8, 1985 (2019), publisher: De Gruyter.
- Isoya et al. [1990] J. Isoya, H. Kanda, and Y. Uchida, EPR studies of interstitial Ni centers in synthetic diamond crystals, Physical Review B 42, 9843 (1990).
- Kato et al. [2023] K. Kato, R. Sasaki, K. Matsuura, K. Usami, and Y. Nakamura, High-cooperativity cavity magnon-polariton using a high-Q dielectric resonator, Journal of Applied Physics 134, 083901 (2023).
- Hamamoto et al. [2024] T. Hamamoto, A. Bhunia, R. K. Bhattacharya, H. Takahashi, and Y. Kubo, Dielectric microwave resonator with large optical apertures for spin-based quantum devices, Applied Physics Letters 124, 234001 (2024).
- Faramarzi et al. [2024] F. Faramarzi, R. Stephenson, S. Sypkens, B. H. Eom, H. LeDuc, and P. Day, A 4–8 GHz kinetic inductance traveling-wave parametric amplifier using four-wave mixing with near quantum-limited noise performance, APL Quantum 1, 036107 (2024).
- Parker et al. [2022] D. J. Parker, M. Savytskyi, W. Vine, A. Laucht, T. Duty, A. Morello, A. L. Grimsmo, and J. J. Pla, Degenerate Parametric Amplification via Three-Wave Mixing Using Kinetic Inductance, Physical Review Applied 17, 034064 (2022), publisher: American Physical Society.
- Probst et al. [2015] S. Probst, F. B. Song, P. A. Bushev, A. V. Ustinov, and M. Weides, Efficient and robust analysis of complex scattering data under noise in microwave resonators, Review of Scientific Instruments 86, 024706 (2015).
- Walls and Milburn [2008] D. Walls and G. J. Milburn, eds., Quantum Optics (Springer, Berlin, Heidelberg, 2008).
- Kubo et al. [2010] Y. Kubo, F. R. Ong, P. Bertet, D. Vion, V. Jacques, D. Zheng, A. Dréau, J.-F. Roch, A. Auffeves, F. Jelezko, J. Wrachtrup, M. F. Barthe, P. Bergonzo, and D. Esteve, Strong Coupling of a Spin Ensemble to a Superconducting Resonator, Physical Review Letters 105, 140502 (2010).
- Choi and Englund [2023] H. Choi and D. Englund, Ultrastrong magnetic light-matter interaction with cavity mode engineering, Communications Physics 6, 1 (2023), number: 1 Publisher: Nature Publishing Group.
- Collett and Gardiner [1984] M. J. Collett and C. W. Gardiner, Squeezing of intracavity and traveling-wave light fields produced in parametric amplification, Physical Review A 30, 1386 (1984), publisher: American Physical Society.
- Diniz et al. [2011] I. Diniz, S. Portolan, R. Ferreira, J. M. Gérard, P. Bertet, and A. Auffèves, Strongly coupling a cavity to inhomogeneous ensembles of emitters: Potential for long-lived solid-state quantum memories, Physical Review A 84, 10.1103/PhysRevA.84.063810 (2011).
- Kurucz et al. [2011] Z. Kurucz, J. H. Wesenberg, and K. Mølmer, Spectroscopic properties of inhomogeneously broadened spin ensembles in a cavity, Physical Review A 83, 10.1103/PhysRevA.83.053852 (2011).
- Julsgaard and Mølmer [2012] B. Julsgaard and K. Mølmer, Dynamical evolution of an inverted spin ensemble in a cavity: Inhomogeneous broadening as a stabilizing mechanism, Physical Review A 86, 063810 (2012).
- Carmichael [2013] H. J. Carmichael, Statistical Methods in Quantum Optics 1: Master Equations and Fokker-Planck Equations (Springer Science & Business Media, 2013) google-Books-ID: 5hv2CAAAQBAJ.
- Schweiger and Jeschke [2001] A. Schweiger and G. Jeschke, Principles of Pulse Electron Paramagnetic Resonance (Oxford University Press, Oxford, New York, 2001).
- Iakoubovskii and Stesmans [2002] K. Iakoubovskii and A. Stesmans, Dominant paramagnetic centers in 17 O -implanted diamond, Physical Review B 66, 045406 (2002).
- Meesala et al. [2018] S. Meesala, Y.-I. Sohn, B. Pingault, L. Shao, H. A. Atikian, J. Holzgrafe, M. Gündoğan, C. Stavrakas, A. Sipahigil, C. Chia, R. Evans, M. J. Burek, M. Zhang, L. Wu, J. L. Pacheco, J. Abraham, E. Bielejec, M. D. Lukin, M. Atatüre, and M. Lončar, Strain engineering of the silicon-vacancy center in diamond, Physical Review B 97, 205444 (2018), publisher: American Physical Society.
- Kurokawa et al. [2024] H. Kurokawa, K. Wakamatsu, S. Nakazato, T. Makino, H. Kato, Y. Sekiguchi, and H. Kosaka, Coherent electric field control of orbital state of a neutral nitrogen-vacancy center, Nature Communications 15, 4039 (2024), publisher: Nature Publishing Group.
- Liu et al. [2021] A. Liu, S. T. Cundiff, D. B. Almeida, and R. Ulbricht, Spectral broadening and ultrafast dynamics of a nitrogen-vacancy center ensemble in diamond, Materials for Quantum Technology 1, 025002 (2021), publisher: IOP Publishing.
- Acosta et al. [2013] V. M. Acosta, K. Jensen, C. Santori, D. Budker, and R. G. Beausoleil, Electromagnetically Induced Transparency in a Diamond Spin Ensemble Enables All-Optical Electromagnetic Field Sensing, Physical Review Letters 110, 213605 (2013).
Supplementary Materials for
A near-quantum-limited diamond maser amplifier operating at millikelvin temperatures
I Materials and Methods
I.1 Sample and device
In this section, we describe the diamond sample and the microwave resonator device used in this work.

I.1.1 Diamond sample
The diamond sample used in this work is identical to those described in previous works (Refs. [17, 16]). It was synthesized by the temperature gradient method under high-pressure and high-temperature (HPHT) conditions of and using an Fe-Ni-Co alloy as the metal solvent and natural diamond grains as the carbon source. A plate obtained by laser-cutting and polishing was irradiated with electrons at a fluence of at room temperature, followed by annealing at for 2 hours in vacuum. Subsequently, it was further irradiated with electrons at to a fluence of and annealed at for 2 hours in vacuum. The final dimensions of the diamond () used in this study were obtained from the central part of the plate via laser-cutting and polishing. Following electron irradiation and annealing, the concentrations of P1 centers and negatively-charged nitrogen-vacancy (NV-) centers were measured to be approximately and , respectively.
I.1.2 Loop-gap resonator
The microwave resonator used in this work is a loop-gap type [16]. We determined its internal and external quality factors using the circle-fit method [30] after subtracting background amplitude oscillations to achieve a flat baseline. The results are shown in Fig. S1 for both and . The internal quality factor consistently stays around . The resonator frequency, , is very sensitive to the distances of the two gaps. Consequently, after each disassembly and reassembly of the resonator and its enclosure to adjust , varied over to upon each cooldown.
I.2 Cryogenic and room temperature setups
In this section, we describe the wiring and setup both inside the dilution refrigerator and at room temperature.


I.2.1 Wiring configuration in the dilution refrigerator
The detailed wiring and configuration inside the dilution refrigerator (Bluefors, LD-400) are illustrated in Fig. S2. Cryogenic attenuators (XMA, 2082-624X-CRYO series), represented as gray round rectangles, are inserted in each input line at every temperature stage to attenuate thermal noise from the microwave tone injected from room temperature. The heavily attenuated “Probe” line and moderately attenuated “Pump” line are combined at a directional coupler (Marki Microwave, C20-0226) on the mixing chamber stage. The labels “LPF” and “IRF” represent low-pass filters (Marki Microwave, FLP-0750) and homemade infrared filters using iron-loaded epoxy Eccosorb CR110, respectively. A HEMT (High Electron Mobility Transistor) amplifier (Low Noise Factory, LNA-LNC4_8C) was installed on the plate.
The calibrated noise source [18] is used for the noise temperature measurements. While the noise temperature is heated maximally up to , the base temperature of the mixing chamber plate slightly increases up to .
Two microwave switches (Radiall, R573423600) were used to check the total attenuation of the coaxial cables to and from the noise source, which turned out to be negligible. The microwave switch (Analog Devices Inc., HMC547ALP3E), depicted inside the dashed box in Fig. S2, was installed and used only for pulse electron spin resonance (pulse-ESR) measurements to protect the HEMT. Further details are described in the Supplementary Text.
I.2.2 Pump cancellation
The probe tone was fed into the fridge through the line labeled “Probe”, while the pump tone was sent through the line “Pump”. To minimize the impact of the pump tone on the HEMT and the following receiver circuits, a cancellation microwave tone was introduced through a dedicated microwave input line labeled “Cancellation” (see also Fig. S3). When operating the maser amplifier, a microwave tone was generated by a signal generator (PSG; Keysight Technologies, E8267D) for both pump and cancellation. This tone was then split by a power splitter (Mini-Circuits, ZC2PD-01263-S+), as shown in Fig. S3A. On one branch, the cancellation tone was adjusted using a voltage-controlled phase shifter (Qorvo, CMD297P34) and a voltage-controlled variable attenuator (Pulsar Microwave, AAT-22-479A/7S). The amplitude and phase of this cancellation tone were adjusted to destructively interfere with the reflected pump tone through a directional coupler placed after the two circulators (Quinstar, QCY-G040082AS), resulting in suppression of the pump power by at least .
I.2.3 Setup for maser gain and noise temperature measurement
The reflection coefficient of the resonator was measured using a vector network analyzer (VNA; Keysight, N5230C). The data presented in Figs. 2C and 3A, B, and D in the main text were measured at a power of , corresponding to a mean intra-resonator photon number of approximately , which is sufficiently lower than the number of inverted spins, , ensuring minimal impact on the spin saturation. Although this power level almost reaches the typical saturation power of traveling-wave parametric amplifiers [2], it remains well below the saturation power of our maser amplifier as discussed in Fig. 4C in the main text.
Noise power spectra were measured by a spectrum analyzer (SPA; Keysight Technologies; N9010A). The output and input ports of the VNA are connected to the Probe and Output ports of the fridge via directional couplers (input: Pulsar Microwave, CS20-09-436/9; output: Marki, C10-0226). A signal generator (SG, Keysight Technologies, E8247C) generates a microwave probe tone during the signal-to-noise ratio (SNR) measurements, as presented in Fig. 4A.
I.2.4 Setup for compression measurement
I.2.5 Setup for pulse electron spin resonance measurements
In the pulse ESR measurement detailed in the Supplementary Text, we use an FPGA-based integrated measurement system (Quantum Machines, OPX+) to generate pulse envelopes, control microwave switches, and detect spin echo signals. Gaussian-shaped pulse envelope signals, modulated at , are generated by the OPX arbitrary waveform generator (AWG). We use the single sideband modulation technique, where a baseband microwave tone (“LO” in Fig. S3B) is mixed with the pulse envelopes with an internal IQ mixer of the PSG. These pulses are subsequently amplified by a high-power amplifier (Cree, CMPA601C025F) and fed to the resonator through the “Pump” line. To minimize microwave leakage noise from the amplifier, a semiconductor-based microwave switch (Analog Devices Inc., ADRF5019-EVALZ) is installed immediately after the amplifier’s output. Similarly, another semiconductor-based microwave switch (Analog Devices Inc., HMC547ALP3E) is installed to protect the HEMT amplifier from the pulses.
For echo signal detection, the signals from the output port of the refrigerator are amplified by a series of room-temperature amplifiers (B&Z Technologies, BZ-02000800-090826-182020, and Mini-Circuits, ZVA-183+) and then down-converted using an IQ mixer (Marki, IQ-0307L). The local oscillator (LO) tones are extracted from a portion of the PSG’s locking signal (option HCC). After demodulation, the signals are passed through band-pass filters (Mini-Circuits, SBP-101+), and then amplified by a preamplifier (SRS, SR445A). Finally, they are detected by the OPX’s digitizer, which processes the signals digitally for heterodyne detection.
I.3 Noise temperature measurement
In this section, we detail the methodology used to measure the noise temperature of our maser amplifier using an impedance-matched temperature-variable noise source.
I.3.1 Y-factor method using a noise source
We characterized the noise temperature of the maser amplifier by using a procedure based on the Y-factor method with a calibrated noise source [18]. The noise source is weakly thermalized to the cold plate at about , which allows for quasi-independent temperature control using an attached heater and thermometer. This setup enables precise control over the input noise temperature into the maser amplifier device.
Thermal noise power is given by , where , , and are the Boltzmann constant, the noise temperature, and the measurement bandwidth, respectively. When a signal with a noise temperature of is fed into an amplifier, the output power is given by
(S1) |
where and are the power gain and noise temperature of the amplifier, respectively. In our experiment, without the maser amplifier, can thus be expressed as:
(S2) |
where and represent the gain and noise temperature of the HEMT amplifier, respectively. The additional noise contribution from the final room temperature amplifier, including cable attenuation from the HEMT to the room temperature setup, is denoted as . Equation (S2) can be approximated to Eq. (1) in the main text, considering that and .
I.3.2 Impact of insertion losses and attenuation on noise temperature
In our setup, a directional coupler and a circulator, as well as lossy non-superconducting cables, are placed between the noise source and the maser amplifier. Their insertion losses and attenuations must, therefore, be taken into account in the evaluation of the input noise temperature , as they attenuate the noise photons from the noise source. To model this effect, we employ the beam-splitter model [31]. If the -th component in the transmission line has an insertion loss , the number of noise photons can be expressed as
(S4) |
where is the number of input noise photons of the -th component, and is the thermal noise photons corresponding to the physical temperature of the component. All the presented values of are corrected using Eq. (S4) based on the measurements detailed in the Supplementary Texts.
I.4 Extracting spin information through resonator
In this section, we describe how we extracted the spectral information of the spin ensemble through the resonator’s reflection coefficient using the input-output theory. This enables us to determine the corresponding number of spins, as shown in Fig. 3A in the main text.
I.4.1 Single Spin Coupling Constant
The single-spin coupling constant , which represents the interaction strength between a single P1 center and a single microwave photon in a resonator, is expressed as [32]
(S5) |
where and denotes the gyromagnetic ratio of the electron spin and the magnetic vacuum fluctuation of the resonator, respectively, and is the spin vector operator with and are the P1 center’s electron spin eigenstates. The magnetic vacuum fluctuation is given by
(S6) |
where , , and are the vacuum permeability, the resonator frequency, and the effective mode volume of the resonator, respectively. is defined as the ratio of the integral of the magnetic field energy density over the whole resonator mode volume to the maximum magnetic energy density [33]
(S7) |
where is the magnetic field in the resonator, maximized at the position . Using a finite element simulation software, COMSOL Multiphysics, the effective resonator mode volume of our loop-gap resonator is estimated to be approximately . Therefore, the single-spin coupling constant of the P1 center with is obtained to be .
I.4.2 Input-output theory
To determine the number of inverted or non-inverted spins through the resonator reflection , we employed the input-output theory [34] for a coupled resonator-spin ensemble system with a coherent probe tone [17, 10]. The Hamiltonian can be expressed using the Tavis-Cummings model with Holstein-Primakoff approximation [35, 36, 37] as follows:
(S8) |
where () and () denote the annihilation and creation operators for the resonator mode (-th spin), respectively. In addition, represents the number of spins, and is the single spin coupling constant between the resonator and the -th spin. and are the external and internal loss rates of the resonator, with () representing the external (internal) quality factor.
I.4.3 The spin function
Following the standard method for deriving the reflection coefficient [34], we obtain
(S9) |
where , called the “spin function”, includes all information about the spin ensemble. As will be discussed in the following section, is linked to the spin susceptibility [17] and is defined as shown in Refs. [35, 36, 37],
(S10) |
with representing the ensemble coupling constant between the resonator and the spin ensemble. The decay rate of individual spins is denoted by , while and represent the center frequency and the half-width at half-maximum (HWHM) of the spin ensemble, respectively, assumed to follow a Lorentzian distribution here. The sign in front of in Eq. (S9) represents the state of the spin ensemble, i.e., a positive sign () corresponds to an inverted spin ensemble, whereas a negative sign () represents a non-inverted spin ensemble.
We modify to incorporate the spin polarization , where represents the population difference between the lower and upper levels of the state, and is the total number of spins in the transition. In a system with a population difference , the ensemble coupling constant is replaced by the effective coupling constant , defined as . Consequently, is reformulated in terms of as
(S11) |
where represents the “conventional” collective coupling constant, assuming all spins are polarized into the lower energy state. The sign of reflects the sign of ; it becomes negative in the case of population inversion and positive in a non-inverted state. To ensure consistency with Eq. (S9), we rewrite the reflection coefficient as follows
(S12) |
I.4.4 Measurement of population differences
The function can be expressed in terms of the reflection coefficient by rearranging Eq. (S12):
(S13) |
which allows to be directly determined from the experimentally obtained data. Fig. S4 shows such examples converted from experimentally obtained data and fits using the imaginary parts of Eq. (S11).
In an ideal spin ensemble, the imaginary part of exhibits a Lorentzian profile, enabling straightforward determination of the population difference by fitting the measured spectrum using the function:
(S14) |
From this function and the imaginary part of Eq. (S11), the population difference can be extracted as .

However, significant deviations from the ideal Lorentzian profile are observed under certain experimental conditions. Immediately after initiating the pump, and before reaching the steady state, the spectrum becomes distorted due to the rapid inversion of the spins near the center of the distribution, while those farther from the center remain non-inverted. Furthermore, even in the steady state, intermediate pump powers lead to partial inversion of the spin ensemble, resulting in non-uniform inversion and persistent spectral distortion. This behavior is illustrated in Fig. S4. Panel A shows spectra at selected time points extracted from the time evolution in Fig. 3A of the main text. Among these, the spectra at 146.6 and 326.7 seconds clearly exhibit transient distortions following pump initiation. Panel B shows spectra at various pump powers extracted from the gain profiles in Fig. 3B. Notably, the spectrum at a pump power of exhibits pronounced steady-state distortion due to partial inversion.
To accurately account for these spectral conditions, we employ a two-Lorentzian fitting model in which the spectrum is represented as a superposition of two separate Lorentzian contributions; one representing the inverted upper-level spins and the other the lower-level spins:
(S15) |
The results of fits using this model are illustrated in Fig. S4. From this fitting model, the population difference can be described as
(S16) |
I.4.5 Maser gain
The peak maser gain is defined by the absolute square of the reflection coefficient at the resonance, where . According to Eq. (S12), this relationship can be expressed as
(S17) |
where we defined a new parameter, [10, 37], as the “spin transition rate”. This parameter quantifies the effective transition rate of the spin ensemble to the resonator mode, i.e., the rate at which spins undergo upwards (downwards) transitions per unit time through the stimulated emission (absorption) with a rate . Moreover, is directly correlated with the “magnetic quality factor” [8, 12], which characterizes a spin ensemble system as a maser gain medium:
(S18) |
where is the magnetic field filling factor and is the imaginary part of the spin susceptibility at resonance, defined as [17].
I.5 Modeling spin dynamics
In this section, we describe the semi-classical rate equations we used to analyze the spin system’s dynamics.
I.5.1 Semi-classical Rate Equations
We modeled the dynamics of our diamond maser amplifier system as detailed below and solved them using semi-classical rate equations [19]. Although quantum Langevin equations would yield equivalent results [9, 10], our lack of expertise with that numerical treatment led us to opt for a semi-classical approach. The system under consideration, as depicted in Fig. S5A, consists of eight energy levels, which include two levels of the “fast-relaxing spins” and the three two-level systems associated with the P1 centers nuclear spin states, P1+, P10, and P1-. This configuration leads to eight rate equations, in addition to a photon number rate equation, to describe the population dynamics and photon numbers within the system. However, due to the computational complexity of solving nine equations, we simplified the model to seven equations by combining the two levels of with the fast-relaxing spins. The resulting set of seven rate equations, including the photon number rate equation, are represented as follows [19, 38]:
(S19) | ||||
where represents the number of spins of each level, , , and are the relaxation rates of each transition, is the four-spin cross-relaxation rate, and is the stimulated emission rate. The rate equation for the intra-resonator photon number is:
(S20) |

I.5.2 Cross-relaxation rate
Parameter | (Hz) | (Hz) | (Hz) | (Hz) |
---|---|---|---|---|
Value |
(Hz) | (Hz) | (Hz) | (Hz) | |
---|---|---|---|---|
The cross-relaxation rate in our system cannot be directly measured due to the presence of the “fast-relaxing spins”, contrasting with the previous reports [19, 20, 21]. To estimate this rate indirectly, we compared computational results from the rate equations with experimental results obtained under various pump powers, as depicted in Fig. S5B and explained in the next subsection. We determined to be within the range of to , with the parameters listed in Tables S1 and S2. While previously reported cross-relaxation rates for highly nitrogen-doped diamonds typically range from to [19, 20], theoretical models of the four-spin interactions imply that [20], where denotes the spin density. This supports our findings of , as the P1 center density in our diamond sample of is considerably lower than the in those previous studies.
I.5.3 Calculation of pump rate
Spins absorb incoming microwave pump tones and are excited to the upper state, a process governed by the same underlying physics as stimulated emission. Therefore, the photon absorption rate, i.e., the pump rate , is calculated as a product of the stimulated emission rate and the number of incoming photons, as described in Eqs. (7.7) and (7.8) on p.260 in Ref. [38]:
(S21) |
where is the pump frequency, is the average spin resonance frequency of P10, and is the number of input photons. The expression on the right-hand side is the rigorous stimulated emission rate with detuning between and .
From the input-output theory, with a coherent pump tone applied to a single-port resonator is expressed as
(S22) |
where denotes the input pump power, and is the total loss rate of the resonator. Therefore, the pump rate can be expressed as a function of as follows;
(S23) |
During maser operation, the pump frequency is tuned to match the resonance frequency of the P10 spins, i.e., , but it is off-resonance with the resonator, thus . Figure S5B displays the population difference in the P1+ transition, , plotted against the calculated pump rates . This population difference is extracted using the spin function , as discussed in the above section. The plot also includes overlayed simulation results from the rate equations Eqs. (S19) and (S20).
I.5.4 Self-oscillation threshold
If the photons produced by stimulated emission exceed the resonator losses, the system enters the self-oscillating or free-running maser regime. In this regime, the maser is capable of emitting microwave photons into the resonator without the need for an externally applied probe tone. The threshold population difference can be calculated from the photon number rate equation given in Eq. (S20) [38]. At the threshold, the two terms in the equation are balanced, hence,
(S24) |
where is the number of spins in the state 4 at the threshold and is the photon number in the steady state in the free-running maser regime. Rearranging this equation gives
(S25) |
The second term can often be neglected since typically . Therefore,
(S26) |
In the case of , is approximately , while for , it is approximately . These thresholds are shown as the red dot-dashed lines in the insets of Fig. 3A and C in the main text.
II Supplementary Text

II.1 Echo-detected field-sweep spectroscopy
To characterize the impurity spins in our diamond, specifically the “fast-relaxing spins,” we conducted an echo-detected field-sweep (EDFS) measurement using the setup depicted in Fig. S3B, with the pulse sequence illustrated in Fig. S6A. A practical challenge arises from the “long relaxing spins”, namely the P1 and NV- centers, which exhibit very long relaxation times ranging from approximately to seconds. Additionally, the substantial quantity of these impurity spins may hinder the signals from the “fast-relaxing spins”, which turned out to be about two orders of magnitude less abundant. To mitigate this, we first saturated the P1 and NV- centers using a 200-ms-long saturation pulse with frequency modulation of centered at the resonator frequency of in this cooldown. This allowed us to primarily detect the spin echo signals from the “fast-relaxing spins”. This approach also enabled optimal adjustment of the room temperature amplifier chain to utilize the full dynamic range of the digitizer. After a waiting time of , the echo signal was detected using a Hahn-echo sequence consisting of and pulses.
The results of the EDFS measurement are shown in Fig. S6B. Despite the saturation pulse, several peaks associated with the P1 and NV- centers remain visible. This is attributed to imperfections in the saturation pulse and partial population recovery during the wait time. Nevertheless, a distinct “broad shoulder” feature is observed, extending from around the P1- peak at approximately to around . We fitted the data with a sum of nine Lorentzian functions. These components are also indicated by short arrows at the top of Fig. S6B. The broad signal stemming from the “fast-relaxing spins” is represented by a solid red curve.
II.2 Characterization of “fast-relaxing spins”

We measured the longitudinal relaxation times of the “fast-relaxing spins” under constant magnetic fields ranging from to using the saturation recovery pulse sequence [39]. The results are shown in Fig. S6C and D. Near the transition of P1-, the relaxation curves exhibit bi-exponential decays, as presented in the upper panel of Fig. S6C. In contrast, in the fields higher than , the relaxation curves can be fitted by a single exponential decay (bottom panel of Fig. S6C). This is attributed to the cross-relaxation effects with the P1- transition and their 13C hyperfine-coupled transition.
As noted in the main text, we have not been able to rigorously identify the specific spin species constituting the “fast-relaxing spins.” Nevertheless, the evidence presented below suggests that these may be neutrally-charged nitrogen-vacancy (NV0) centers. First, we measured of P1 centers in a type Ib yellow diamond, an as-grown HPHT sample purchased from Sumitomo Electric Industries, Ltd., containing approximately to of P1 centers, at . The results, displayed in Fig. S7B, show spin relaxation times exceeding seconds in the short , about four or five times longer than those measured in the sample used for the maser experiments as presented in Fig. S7A. Similarly, the longer also show and seconds. These imply that the electron irradiation, i.e., the introduction of vacancies into our diamond crystal, likely created the “fast-relaxing spins” that exhibit the broad spin distribution (Fig. S6B), subsequently impacting the transitions of P1 centers.
Vacancy clusters [40] initially appeared to be possible candidates. However, these defects typically exhibit spin transitions with much narrower linewidths (significantly less than ) at temperatures below [40], which stands in stark contrast to the broad transition observed for the fast-relaxing spins in our diamond crystal. On the other hand, similar to negatively-charged silicon-vacancy (SiV-) centers [41], the spins of NV0 centers are sensitive to crystal strain [23, 42]. Such strain becomes significant in diamond crystals containing defect concentrations exceeding [43, 44], which is indeed the case for the diamond crystal used in our experiments.
II.3 Noise temperature of baseline

We measured the noise temperature at several frequencies around the maser peak gain, both with the pump on and off, as shown in Fig. S8. First, we measured the noise temperature with the pump off, namely, . To minimize the absorption effect by the spins, which is equivalent to having additional attenuation along the line, the magnetic field was detuned from the resonance of P1+.
Under these conditions, as seen in Fig. S8, the noise temperatures are measured to be between and , in reasonable agreement with the values reported in the HEMT’s datasheet. The oscillations are mostly attributed to small impedance mismatches among the components along the measurement chain from up to the HEMT amplifier. At the resonance frequency , it reaches , which is used as the HEMT’s noise temperature in the main text. During this measurement with the pump off, we disconnected both the pump and cancellation lines and terminated at room temperature to minimize the noise from the room-temperature setup, especially from the amplifier in the cancellation line (Fig. S3).
When the pump and cancellation tones are turned on and tuned for maser amplifier’s gain of , the noise baseline increases by approximately over the entire frequency range (except near , i.e., the maser amplifier frequency), most likely due to noise coming from the room-temperature amplifier connected to the “Cancellation” line. The significant increase of the noise around is due to the pump.
II.4 Detailed line attenuation characterization

In order to accurately estimate the power of the pump and probe tones arriving at the input of the maser amplifier device, we independently characterized the attenuation of both the probe and pump lines in a separate cooldown. To this end, we measured the transmission coefficients over the “symmetric lines”, i.e, pairs of lines with identical configurations, including coaxial cable materials, lengths, and attenuators at each temperature stage, as illustrated in Fig. S9. These symmetric lines were interconnected by a short, low-loss coaxial cable at the mixing chamber stage.
We also characterized the attenuation of the cables and components in the Probe line at the mixing chamber stage, specifically from the infrared filter through the circulator, noise source, directional coupler, and the final circulator to the input of the maser device (see also Fig. S2). This was performed using two mechanical microwave switches (Radiall, R573423600), as depicted in Fig. S9. Four measurement paths (Ch. 1–Ch. 4) were established for this purpose. Channels 1 and 2 correspond to the paths from the noise source to the maser amplifier device and from the infrared filter to the noise source, respectively. Channel 3 was used to isolate the attenuation introduced by the components measured in Ch. 1 and Ch. 2 by using a low-loss flexible coaxial cable. Channel 4 was used to estimate the attenuation of the low-loss flexible coaxial cables used to connect the components by comparing with Ch. 3. The results of these measurements are summarized in Table S3. Note that the insertion loss of the mechanical switches must also be taken into account in the calculation of Eq. (S4) for the noise temperature measurement; here we relied on the value on the datasheet of .
Line | Attenuation value (dB) |
---|---|
Probe | 53.3 |
Pump, Cancellation | 27.6 |
Channel 1 | 1.26 |
Channel 2 | 1.50 |